A Comparative Analysis of Monte Carlo and Collapsed Cone Dose Calculation Algorithms for Monaco 3D Treatment Plans

Download pdf | Full Screen View

Date Issued:
2022
Abstract/Description:
The Monaco treatment planning system offers three different dose calculation algorithms for use in calculating 3D treatment plans. These include Monte Carlo (MC), Collapsed Cone (CC) and the pencil beam algorithms. The aim of this study is an in-depth analysis of Monte Carlo and Collapsed Cone dose calculation methods to find the optimal parameters for clinical use for both algorithms. An end-to-end phantom with inhomogeneities was scanned and the DICOM images were imported into Monaco for contouring and planning. Treatment plans were then created in Monaco for both MC and CC using different permutations of variables for approximately 400 plans. These variables include CT Slice thickness, grid size, statistical uncertainty, and beam energy. Following planning the end-to-end phantom was then irradiated on an Elekta Linac and plans for each beam energy were created. Clinical beam data was then compared to the computed plans for each dose calculation method.
Title: A Comparative Analysis of Monte Carlo and Collapsed Cone Dose Calculation Algorithms for Monaco 3D Treatment Plans.
69 views
29 downloads
Name(s): Pereira, Shakeel , author
Kyriacou, Andreas , Thesis advisor
Florida Atlantic University, Degree grantor
Department of Physics
Charles E. Schmidt College of Science
Type of Resource: text
Genre: Electronic Thesis Or Dissertation
Date Created: 2022
Date Issued: 2022
Publisher: Florida Atlantic University
Place of Publication: Boca Raton, Fla.
Physical Form: application/pdf
Extent: 84 p.
Language(s): English
Abstract/Description: The Monaco treatment planning system offers three different dose calculation algorithms for use in calculating 3D treatment plans. These include Monte Carlo (MC), Collapsed Cone (CC) and the pencil beam algorithms. The aim of this study is an in-depth analysis of Monte Carlo and Collapsed Cone dose calculation methods to find the optimal parameters for clinical use for both algorithms. An end-to-end phantom with inhomogeneities was scanned and the DICOM images were imported into Monaco for contouring and planning. Treatment plans were then created in Monaco for both MC and CC using different permutations of variables for approximately 400 plans. These variables include CT Slice thickness, grid size, statistical uncertainty, and beam energy. Following planning the end-to-end phantom was then irradiated on an Elekta Linac and plans for each beam energy were created. Clinical beam data was then compared to the computed plans for each dose calculation method.
Identifier: FA00014019 (IID)
Degree granted: Thesis (PMS)--Florida Atlantic University, 2022.
Collection: FAU Electronic Theses and Dissertations Collection
Note(s): Includes bibliography.
Subject(s): Radiotherapy
Radiotherapy Dosage
Radiation dosimetry
Persistent Link to This Record: http://purl.flvc.org/fau/fd/FA00014019
Use and Reproduction: Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
Use and Reproduction: http://rightsstatements.org/vocab/InC/1.0/
Host Institution: FAU
Is Part of Series: Florida Atlantic University Digital Library Collections.