Efficacy of the virtual cone method using fixed small multi-leaf collimator field for stereotactic radiosurgery
- Date Issued:
- 2022
- Abstract/Description:
- Dosimetric uncertainty in very small (< 2 x 2 cm2) photon fields is notably higher that has created research questions when using small-field virtual cone with variable multileaf collimator (MLC) fields. We evaluate the efficacy of the virtual cone with a fixed MLC field for stereotactic radiosurgery (SRS) of small targets such as trigeminal neuralgia. We employed a virtual cone technique with a fixed field geometry, called fixed virtual cone (fVC), for small target radiosurgery using the EDGE (Varian Medical Systems, Palo Alto, CA) linac. The fVC is characterized by 0.5 cm x 0.5 cm high-definition MLC field of 10 MV flattening filter-free (FFF) beam defined at 100 cm SAD, while jaws are positioned at 1.5 cm x 1.5 cm. A spherical dose distribution equivalent to 5 mm cone was generated by using 10–14 non-coplanar partial arcs. The dosimetric accuracy of this technique was validated using the SRS MapCHECK (Sun Nuclear Corporation, FL) and the EBT3 (Ashland Inc., NJ) film based on absolute dose measurements. For the quality assurance (QA), 10 treatment plans for trigeminal neuralgia consisting of various arc fields at different collimator angles were analyzed retrospectively using 6 MV and 10 MV FFF beams, including the field-by-field study (n = 130 fields). Dose outputs were compared between the SRS MapCHECK measurements and Eclipse treatment planning system (TPS) with Acuros XB algorithm (version 16.1). In addition, important clinical parameters of 15 cases treated for trigeminal neuralgia were evaluated for the clinical performance. Moreover, dosimetric (field output factors, dose/MU) uncertainties considering a minute (± 0.5–1.0 mm) leaf shift in the field defining fVC, were examined from the TPS, SRS diode (PTW 60018) measurements, and Monte Carlo (MC) simulations.
Title: | Efficacy of the virtual cone method using fixed small multi-leaf collimator field for stereotactic radiosurgery. |
84 views
32 downloads |
---|---|---|
Name(s): |
Neupane, Taindra, author Leventouri, Theodora, Thesis advisor Shang, Charles, Thesis advisor Florida Atlantic University, Degree grantor Department of Physics Charles E. Schmidt College of Science |
|
Type of Resource: | text | |
Genre: | Electronic Thesis Or Dissertation | |
Date Created: | 2022 | |
Date Issued: | 2022 | |
Publisher: | Florida Atlantic University | |
Place of Publication: | Boca Raton, Fla. | |
Physical Form: | application/pdf | |
Extent: | 128 p. | |
Language(s): | English | |
Abstract/Description: | Dosimetric uncertainty in very small (< 2 x 2 cm2) photon fields is notably higher that has created research questions when using small-field virtual cone with variable multileaf collimator (MLC) fields. We evaluate the efficacy of the virtual cone with a fixed MLC field for stereotactic radiosurgery (SRS) of small targets such as trigeminal neuralgia. We employed a virtual cone technique with a fixed field geometry, called fixed virtual cone (fVC), for small target radiosurgery using the EDGE (Varian Medical Systems, Palo Alto, CA) linac. The fVC is characterized by 0.5 cm x 0.5 cm high-definition MLC field of 10 MV flattening filter-free (FFF) beam defined at 100 cm SAD, while jaws are positioned at 1.5 cm x 1.5 cm. A spherical dose distribution equivalent to 5 mm cone was generated by using 10–14 non-coplanar partial arcs. The dosimetric accuracy of this technique was validated using the SRS MapCHECK (Sun Nuclear Corporation, FL) and the EBT3 (Ashland Inc., NJ) film based on absolute dose measurements. For the quality assurance (QA), 10 treatment plans for trigeminal neuralgia consisting of various arc fields at different collimator angles were analyzed retrospectively using 6 MV and 10 MV FFF beams, including the field-by-field study (n = 130 fields). Dose outputs were compared between the SRS MapCHECK measurements and Eclipse treatment planning system (TPS) with Acuros XB algorithm (version 16.1). In addition, important clinical parameters of 15 cases treated for trigeminal neuralgia were evaluated for the clinical performance. Moreover, dosimetric (field output factors, dose/MU) uncertainties considering a minute (± 0.5–1.0 mm) leaf shift in the field defining fVC, were examined from the TPS, SRS diode (PTW 60018) measurements, and Monte Carlo (MC) simulations. | |
Identifier: | FA00013958 (IID) | |
Degree granted: | Dissertation (Ph.D.)--Florida Atlantic University, 2022. | |
Collection: | FAU Electronic Theses and Dissertations Collection | |
Note(s): | Includes bibliography. | |
Subject(s): |
Radiation dosimetry Radiosurgery Collimators (Optical instrument) Monte Carlo method |
|
Persistent Link to This Record: | http://purl.flvc.org/fau/fd/FA00013958 | |
Use and Reproduction: | Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder. | |
Use and Reproduction: | http://rightsstatements.org/vocab/InC/1.0/ | |
Host Institution: | FAU | |
Is Part of Series: | Florida Atlantic University Digital Library Collections. |