You are here

Commissioning of 360⁰ Rotational Single Room ProBeam Compact™ (Varian Medical) Pencil Beam Scanning Proton Therapy System

Download pdf | Full Screen View

Date Issued:
2021
Summary:
A clinical commissioning of the first 360 rotational compact Varian ProBeam scanning proton pencil beam (Varian Medical, Palo Alto, CA) system was conducted at the South Florida Proton Therapy Institute (SFPTI). The beam dosimetry and characterizations were the vital section used to verify the consistency of the treatment planning system (TPS) outputs. The integrated depth dose curves were acquired with AP CAX in water phantom utilizing a large PTW Bragg peak chamber; the dose output factors measurements were performed by using IBA PCC05 chamber at 1.5 cm water depth applying a single layer 10×10 cm2 beams and 1.1 RBE offset as recommended in TRS 398 report. Widths of the Bragg peaks ranges (Rb80-Ra80) were from 4.07 cm to 30.51 cm for the energy range 70 MeV to 220 MeV. Beam optics such as spot sizes and spot profiles were acquired in-air by using Logos scintillators with a CCD camera and the result data were from 2.33 mm to for 77 MeV to 9.70 mm for 220 MeV. In different field sizes, a comparison between the dose measured using PTW Semiflex and the AcurosPT estimated dose were performed to study the halo effect. All the measured dosimetric parameters showed that the design specifications were well achieved, and the results are suitable for being used as a part of the clinical commissioning and quality assurance program for treating patients.
Title: Commissioning of 360⁰ Rotational Single Room ProBeam Compact™ (Varian Medical) Pencil Beam Scanning Proton Therapy System.
168 views
94 downloads
Name(s): Fathallah, Shreen Mohamed, author
Shang, Charles, Thesis advisor
Muhammad, Wazir, Thesis advisor
Florida Atlantic University, Degree grantor
Department of Physics
Charles E. Schmidt College of Science
Type of Resource: text
Genre: Electronic Thesis Or Dissertation
Date Created: 2021
Date Issued: 2021
Publisher: Florida Atlantic University
Place of Publication: Boca Raton, Fla.
Physical Form: application/pdf
Extent: 80 p.
Language(s): English
Summary: A clinical commissioning of the first 360 rotational compact Varian ProBeam scanning proton pencil beam (Varian Medical, Palo Alto, CA) system was conducted at the South Florida Proton Therapy Institute (SFPTI). The beam dosimetry and characterizations were the vital section used to verify the consistency of the treatment planning system (TPS) outputs. The integrated depth dose curves were acquired with AP CAX in water phantom utilizing a large PTW Bragg peak chamber; the dose output factors measurements were performed by using IBA PCC05 chamber at 1.5 cm water depth applying a single layer 10×10 cm2 beams and 1.1 RBE offset as recommended in TRS 398 report. Widths of the Bragg peaks ranges (Rb80-Ra80) were from 4.07 cm to 30.51 cm for the energy range 70 MeV to 220 MeV. Beam optics such as spot sizes and spot profiles were acquired in-air by using Logos scintillators with a CCD camera and the result data were from 2.33 mm to for 77 MeV to 9.70 mm for 220 MeV. In different field sizes, a comparison between the dose measured using PTW Semiflex and the AcurosPT estimated dose were performed to study the halo effect. All the measured dosimetric parameters showed that the design specifications were well achieved, and the results are suitable for being used as a part of the clinical commissioning and quality assurance program for treating patients.
Identifier: FA00013691 (IID)
Degree granted: Thesis (PMS)--Florida Atlantic University, 2021.
Collection: FAU Electronic Theses and Dissertations Collection
Note(s): Includes bibliography.
Subject(s): Proton Therapy
Dosimetry
Held by: Florida Atlantic University Libraries
Sublocation: Digital Library
Persistent Link to This Record: http://purl.flvc.org/fau/fd/FA00013691
Use and Reproduction: Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
Use and Reproduction: http://rightsstatements.org/vocab/InC/1.0/
Host Institution: FAU
Is Part of Series: Florida Atlantic University Digital Library Collections.