COMMISSIONING AND ACCEPTANCE TESTING OF A TRUEBEAM LINEAR ACCELERATOR

Download pdf | Full Screen View

Date Issued:
2019
Abstract/Description:
Due to the difficulty of a complex commissioning technique for a multi energetic, multi-modality linear accelerator, I perform all the commissioning and acceptance testing for a TrueBeam linear accelerator with 4 megavoltage (MV) energies of which 2 are flattening filter-free (FFF) and 6 electron energies varying from 6 MeV to 20 MeV. A 2 dimensional (2D) water tank was used for scanning all the required field sizes for all the energies. The one dimensional (1D) water tank was used to collect all the output factors for all the photon fields sizes small to medium electron field sizes. For the large electron fields sizes, we had to use the 2D water tank. All the collected data was converted into a file type accepted by the planning system (Eclipse) and subsequently imported there. Treatment plans were generated using multiple forms of planning to verify the viability and quality of the beam data commissioned.
Title: COMMISSIONING AND ACCEPTANCE TESTING OF A TRUEBEAM LINEAR ACCELERATOR.
1959 views
1898 downloads
Name(s): Dumitru, Nicolae, author
Pella, Silvia, Thesis advisor
Florida Atlantic University, Degree grantor
Charles E. Schmidt College of Science
Department of Physics
Type of Resource: text
Genre: Electronic Thesis Or Dissertation
Date Created: 2019
Date Issued: 2019
Publisher: Florida Atlantic University
Place of Publication: Boca Raton, Fla.
Physical Form: application/pdf
Extent: 103 p.
Language(s): English
Abstract/Description: Due to the difficulty of a complex commissioning technique for a multi energetic, multi-modality linear accelerator, I perform all the commissioning and acceptance testing for a TrueBeam linear accelerator with 4 megavoltage (MV) energies of which 2 are flattening filter-free (FFF) and 6 electron energies varying from 6 MeV to 20 MeV. A 2 dimensional (2D) water tank was used for scanning all the required field sizes for all the energies. The one dimensional (1D) water tank was used to collect all the output factors for all the photon fields sizes small to medium electron field sizes. For the large electron fields sizes, we had to use the 2D water tank. All the collected data was converted into a file type accepted by the planning system (Eclipse) and subsequently imported there. Treatment plans were generated using multiple forms of planning to verify the viability and quality of the beam data commissioned.
Identifier: FA00013301 (IID)
Degree granted: Thesis (M.S.)--Florida Atlantic University, 2019.
Collection: FAU Electronic Theses and Dissertations Collection
Note(s): Includes bibliography.
Subject(s): Linear accelerators
Cancer--Treatment
Radiation dosimetry
Held by: Florida Atlantic University Libraries
Sublocation: Digital Library
Persistent Link to This Record: http://purl.flvc.org/fau/fd/FA00013301
Use and Reproduction: Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
Use and Reproduction: http://rightsstatements.org/vocab/InC/1.0/
Host Institution: FAU
Is Part of Series: Florida Atlantic University Digital Library Collections.