You are here
Dosimetric and Radiobiological Comparison of the Effects of High Definition versus Normal Collimation on Treatment Plans for Stereotactic Lung Cancer Radiation Therapy
- Date Issued:
- 2018
- Abstract/Description:
- Stereotactic Body Radiation Therapy (SBRT) is a modern precision radiation therapy to deliver the dose in 1 to 5 fractions with high target dose conformity, and steep dose gradient towards healthy tissues. The dose delivered is influenced by the leaf width of the MLC, especially in case of SBRT. Treatment plans with high definition (HD) MLC having leaf-width 2.5 mm and normal MLC having leaf-width 5 mm, were compared to quantify dosimetric and radiobiological parameters. Dosimetric parameters conformity indices (CI), gradient indices (GI) and heterogeneity indices (HI) were compared. The radiobiological parameters were evaluated by normal tissue complication probability (NTCP) and tumor control probability (TCP) based on the equivalent uniform dose (EUD). The results show that there is dosimetric and radiobiological merit of the HD MLC over the normal MLC. However, the improvement is not consistent with all the plans and thus further research is required prior to conclusion.
Title: | Dosimetric and Radiobiological Comparison of the Effects of High Definition versus Normal Collimation on Treatment Plans for Stereotactic Lung Cancer Radiation Therapy. |
54 views
16 downloads |
---|---|---|
Name(s): |
Pudasaini, Mukunda Prasad, author Pella, Silvia, Thesis advisor Leventouri, Theodora, Thesis advisor Florida Atlantic University, Degree grantor Charles E. Schmidt College of Science Department of Physics |
|
Type of Resource: | text | |
Genre: | Electronic Thesis Or Dissertation | |
Date Created: | 2018 | |
Date Issued: | 2018 | |
Publisher: | Florida Atlantic University | |
Place of Publication: | Boca Raton, Fla. | |
Physical Form: | application/pdf | |
Extent: | 61 p. | |
Language(s): | English | |
Abstract/Description: | Stereotactic Body Radiation Therapy (SBRT) is a modern precision radiation therapy to deliver the dose in 1 to 5 fractions with high target dose conformity, and steep dose gradient towards healthy tissues. The dose delivered is influenced by the leaf width of the MLC, especially in case of SBRT. Treatment plans with high definition (HD) MLC having leaf-width 2.5 mm and normal MLC having leaf-width 5 mm, were compared to quantify dosimetric and radiobiological parameters. Dosimetric parameters conformity indices (CI), gradient indices (GI) and heterogeneity indices (HI) were compared. The radiobiological parameters were evaluated by normal tissue complication probability (NTCP) and tumor control probability (TCP) based on the equivalent uniform dose (EUD). The results show that there is dosimetric and radiobiological merit of the HD MLC over the normal MLC. However, the improvement is not consistent with all the plans and thus further research is required prior to conclusion. | |
Identifier: | FA00013148 (IID) | |
Degree granted: | Thesis (M.S.)--Florida Atlantic University, 2018. | |
Collection: | FAU Electronic Theses and Dissertations Collection | |
Note(s): | Includes bibliography. | |
Subject(s): |
Radiosurgery Lung Neoplasms Radiation dosimetry Radiobiology |
|
Held by: | Florida Atlantic University Libraries | |
Sublocation: | Digital Library | |
Persistent Link to This Record: | http://purl.flvc.org/fau/fd/FA00013148 | |
Use and Reproduction: | Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder. | |
Use and Reproduction: | http://rightsstatements.org/vocab/InC/1.0/ | |
Host Institution: | FAU | |
Is Part of Series: | Florida Atlantic University Digital Library Collections. |