Investigation of Rotational Deviations on Single Fiducial Tumor Tracking with Simulated Respiratory Motion using Synchrony® Respiratory Motion Tracking for Cyberknife® Treatment
- Date Issued:
- 2018
- Abstract/Description:
- It is hypothesized that the uncertainty of the Synchrony® model from the rotation of a geometrically asymmetrical single fiducial shall be non-zero during the motion tracking. To validate this hypothesis, the uncertainty was measured for a Synchrony® model built for a respiratory motion phantom oriented at different yaw angles on a Cyberknife® treatment table. A Mini-ball Cube with three cylindrical GoldMark™ (1mmx5mm Au) numbered fiducials was placed inside a respiratory phantom and used for all tests. The fiducial with the least artifact interference was selected for the motion tracking. A 2cm periodic, longitudinal, linear motion of the Mini-ball cube was executed and tested for yaw rotational angles, 0° – 90°. The test was repeated over 3 nonconsecutive days. The uncertainty increased with the yaw angle with the most noticeable changes seen between20° and 60° yaw, where uncertainty increased from 23.5% to 57.9%. A similar test was performed using a spherical Gold Anchor™ fiducial. The uncertainties found when using the Gold Anchor™ were statistically lower than those found when using the GoldMark™ fiducial for all angles of rotation. For the first time, it is found that Synchrony® model uncertainty depends on fiducial geometry. In addition, this research has shown that tracking target rotation using a single fiducial can be accomplished with the Synchrony® model uncertainty as it is displayed on the treatment console. The results of this research could lead to decreased acute toxicity effects related to multiple fiducials.
Title: | Investigation of Rotational Deviations on Single Fiducial Tumor Tracking with Simulated Respiratory Motion using Synchrony® Respiratory Motion Tracking for Cyberknife® Treatment. |
77 views
22 downloads |
---|---|---|
Name(s): |
Christ, Zachary A., author Shang, Charles, Thesis advisor Leventouri, Theodora, Thesis advisor Florida Atlantic University, Degree grantor Charles E. Schmidt College of Science Department of Physics |
|
Type of Resource: | text | |
Genre: | Electronic Thesis Or Dissertation | |
Date Created: | 2018 | |
Date Issued: | 2018 | |
Publisher: | Florida Atlantic University | |
Place of Publication: | Boca Raton, Fla. | |
Physical Form: | application/pdf | |
Extent: | 45 p. | |
Language(s): | English | |
Abstract/Description: | It is hypothesized that the uncertainty of the Synchrony® model from the rotation of a geometrically asymmetrical single fiducial shall be non-zero during the motion tracking. To validate this hypothesis, the uncertainty was measured for a Synchrony® model built for a respiratory motion phantom oriented at different yaw angles on a Cyberknife® treatment table. A Mini-ball Cube with three cylindrical GoldMark™ (1mmx5mm Au) numbered fiducials was placed inside a respiratory phantom and used for all tests. The fiducial with the least artifact interference was selected for the motion tracking. A 2cm periodic, longitudinal, linear motion of the Mini-ball cube was executed and tested for yaw rotational angles, 0° – 90°. The test was repeated over 3 nonconsecutive days. The uncertainty increased with the yaw angle with the most noticeable changes seen between20° and 60° yaw, where uncertainty increased from 23.5% to 57.9%. A similar test was performed using a spherical Gold Anchor™ fiducial. The uncertainties found when using the Gold Anchor™ were statistically lower than those found when using the GoldMark™ fiducial for all angles of rotation. For the first time, it is found that Synchrony® model uncertainty depends on fiducial geometry. In addition, this research has shown that tracking target rotation using a single fiducial can be accomplished with the Synchrony® model uncertainty as it is displayed on the treatment console. The results of this research could lead to decreased acute toxicity effects related to multiple fiducials. | |
Identifier: | FA00013041 (IID) | |
Degree granted: | Thesis (M.S.)--Florida Atlantic University, 2018. | |
Collection: | FAU Electronic Theses and Dissertations Collection | |
Note(s): | Includes bibliography. | |
Subject(s): |
Fiducial Markers Radiosurgery--Quality control Robotic radiosurgery |
|
Held by: | Florida Atlantic University Libraries | |
Sublocation: | Digital Library | |
Persistent Link to This Record: | http://purl.flvc.org/fau/fd/FA00013041 | |
Use and Reproduction: | Copyright © is held by the author, with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder. | |
Use and Reproduction: | http://rightsstatements.org/vocab/InC/1.0/ | |
Host Institution: | FAU | |
Is Part of Series: | Florida Atlantic University Digital Library Collections. |