You are here

Investigation of Rotational Deviations on Single Fiducial Tumor Tracking with Simulated Respiratory Motion using Synchrony® Respiratory Motion Tracking for Cyberknife® Treatment

Download pdf | Full Screen View

Date Issued:
2018
Abstract/Description:
It is hypothesized that the uncertainty of the Synchrony® model from the rotation of a geometrically asymmetrical single fiducial shall be non-zero during the motion tracking. To validate this hypothesis, the uncertainty was measured for a Synchrony® model built for a respiratory motion phantom oriented at different yaw angles on a Cyberknife® treatment table. A Mini-ball Cube with three cylindrical GoldMark™ (1mmx5mm Au) numbered fiducials was placed inside a respiratory phantom and used for all tests. The fiducial with the least artifact interference was selected for the motion tracking. A 2cm periodic, longitudinal, linear motion of the Mini-ball cube was executed and tested for yaw rotational angles, 0° – 90°. The test was repeated over 3 nonconsecutive days. The uncertainty increased with the yaw angle with the most noticeable changes seen between20° and 60° yaw, where uncertainty increased from 23.5% to 57.9%. A similar test was performed using a spherical Gold Anchor™ fiducial. The uncertainties found when using the Gold Anchor™ were statistically lower than those found when using the GoldMark™ fiducial for all angles of rotation. For the first time, it is found that Synchrony® model uncertainty depends on fiducial geometry. In addition, this research has shown that tracking target rotation using a single fiducial can be accomplished with the Synchrony® model uncertainty as it is displayed on the treatment console. The results of this research could lead to decreased acute toxicity effects related to multiple fiducials.
Title: Investigation of Rotational Deviations on Single Fiducial Tumor Tracking with Simulated Respiratory Motion using Synchrony® Respiratory Motion Tracking for Cyberknife® Treatment.
77 views
22 downloads
Name(s): Christ, Zachary A., author
Shang, Charles, Thesis advisor
Leventouri, Theodora, Thesis advisor
Florida Atlantic University, Degree grantor
Charles E. Schmidt College of Science
Department of Physics
Type of Resource: text
Genre: Electronic Thesis Or Dissertation
Date Created: 2018
Date Issued: 2018
Publisher: Florida Atlantic University
Place of Publication: Boca Raton, Fla.
Physical Form: application/pdf
Extent: 45 p.
Language(s): English
Abstract/Description: It is hypothesized that the uncertainty of the Synchrony® model from the rotation of a geometrically asymmetrical single fiducial shall be non-zero during the motion tracking. To validate this hypothesis, the uncertainty was measured for a Synchrony® model built for a respiratory motion phantom oriented at different yaw angles on a Cyberknife® treatment table. A Mini-ball Cube with three cylindrical GoldMark™ (1mmx5mm Au) numbered fiducials was placed inside a respiratory phantom and used for all tests. The fiducial with the least artifact interference was selected for the motion tracking. A 2cm periodic, longitudinal, linear motion of the Mini-ball cube was executed and tested for yaw rotational angles, 0° – 90°. The test was repeated over 3 nonconsecutive days. The uncertainty increased with the yaw angle with the most noticeable changes seen between20° and 60° yaw, where uncertainty increased from 23.5% to 57.9%. A similar test was performed using a spherical Gold Anchor™ fiducial. The uncertainties found when using the Gold Anchor™ were statistically lower than those found when using the GoldMark™ fiducial for all angles of rotation. For the first time, it is found that Synchrony® model uncertainty depends on fiducial geometry. In addition, this research has shown that tracking target rotation using a single fiducial can be accomplished with the Synchrony® model uncertainty as it is displayed on the treatment console. The results of this research could lead to decreased acute toxicity effects related to multiple fiducials.
Identifier: FA00013041 (IID)
Degree granted: Thesis (M.S.)--Florida Atlantic University, 2018.
Collection: FAU Electronic Theses and Dissertations Collection
Note(s): Includes bibliography.
Subject(s): Fiducial Markers
Radiosurgery--Quality control
Robotic radiosurgery
Held by: Florida Atlantic University Libraries
Sublocation: Digital Library
Persistent Link to This Record: http://purl.flvc.org/fau/fd/FA00013041
Use and Reproduction: Copyright © is held by the author, with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
Use and Reproduction: http://rightsstatements.org/vocab/InC/1.0/
Host Institution: FAU