You are here
Manufacturing of 3D Printed Boluses for Use In Electron Radiation Therapy
- Date Issued:
- 2017
- Summary:
- This research demonstrates that a 3D printed bolus can be customized for electron radiation therapy. Both extruder and powder based printers were used, along with, paraffin wax, super stuff, and H20. The plan dose coverage and conformity for the planning target volume (PTV), was such that the distal side of the PTV was covered by the 90% isodose line. The structure is read, and converted into an STL file. The file is sent to a slicer to print. The object was filled with parafin wax, superstuff or water and sealed. Materials Hounsfield units were analyzed, along with the structure stability. This method is evaluated by scanning the 3D printed bolus. The dose conformity is improved compared to that with no bolus. By generating a patient specific 3D printed bolus there is an in improvement in conformity of the prescription isodose surface while sparing immediately adjacent normal tissues.
Title: | Manufacturing of 3D Printed Boluses for Use In Electron Radiation Therapy. |
221 views
139 downloads |
---|---|---|
Name(s): |
Gibbard, Grant, author Kalantzis, Georgios, Thesis advisor Florida Atlantic University, Degree grantor Charles E. Schmidt College of Science Department of Physics |
|
Type of Resource: | text | |
Genre: | Electronic Thesis Or Dissertation | |
Date Created: | 2017 | |
Date Issued: | 2017 | |
Publisher: | Florida Atlantic University | |
Place of Publication: | Boca Raton, Fla. | |
Physical Form: | application/pdf | |
Extent: | 62 p. | |
Language(s): | English | |
Summary: | This research demonstrates that a 3D printed bolus can be customized for electron radiation therapy. Both extruder and powder based printers were used, along with, paraffin wax, super stuff, and H20. The plan dose coverage and conformity for the planning target volume (PTV), was such that the distal side of the PTV was covered by the 90% isodose line. The structure is read, and converted into an STL file. The file is sent to a slicer to print. The object was filled with parafin wax, superstuff or water and sealed. Materials Hounsfield units were analyzed, along with the structure stability. This method is evaluated by scanning the 3D printed bolus. The dose conformity is improved compared to that with no bolus. By generating a patient specific 3D printed bolus there is an in improvement in conformity of the prescription isodose surface while sparing immediately adjacent normal tissues. | |
Identifier: | FA00005943 (IID) | |
Degree granted: | Thesis (M.S.)--Florida Atlantic University, 2017. | |
Collection: | FAU Electronic Theses and Dissertations Collection | |
Note(s): | Includes bibliography. | |
Subject(s): |
Dissertations, Academic -- Florida Atlantic University Radiotherapy Dosage. Skin--Cancer. Radiotherapy--methods |
|
Held by: | Florida Atlantic University Libraries | |
Sublocation: | Digital Library | |
Persistent Link to This Record: | http://purl.flvc.org/fau/fd/FA00005943 | |
Use and Reproduction: | Copyright © is held by the author, with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder. | |
Use and Reproduction: | http://rightsstatements.org/vocab/InC/1.0/ | |
Host Institution: | FAU | |