MCNP5 Monte Carlo based dosimetry for the Nucletron Iridium-192 high dose-rate brachytherapy source with tissue heterogeneity corrections
- Date Issued:
- 2012
- Summary:
- A Monte Carlo model has been developed using MCNP5 to simulate the Nucletron Ir-192 HDR source in order to investigate the influence of tissue heterogeneities on dose calculations compared to the dose in homogeneous water media, as it is typically calculated by brachytherapy Treatment Planning Systems (TPS). Validity of the simulation was verified in water medium in comparison with peer reviewed results using the dosimetric parameters recommended by AAPM, Task Group-43. The dose-rates in simulated prostate, bladder and rectum were compared to those obtained in the homogeneous water phantom. Based on the resulting dose differences, it is inferred that TPS algorithms for brachytherapy dose calculations overestimate the dose to tissues like prostate and bladder by up to 49%. A clinically relevant dose underestimation of 5.5% to the rectum was also found. We recommend that further investigation using actual patient CT data as input to the Monte Carlo simulation be performed.
Title: | MCNP5 Monte Carlo based dosimetry for the Nucletron Iridium-192 high dose-rate brachytherapy source with tissue heterogeneity corrections. |
248 views
45 downloads |
---|---|---|
Name(s): |
Herrera, Ramses Charles E. Schmidt College of Science Department of Physics |
|
Type of Resource: | text | |
Genre: | Electronic Thesis Or Dissertation | |
Date Issued: | 2012 | |
Publisher: | Florida Atlantic University | |
Physical Form: | electronic | |
Extent: | xvii, 76 p. : ill. (some col.) | |
Language(s): | English | |
Summary: | A Monte Carlo model has been developed using MCNP5 to simulate the Nucletron Ir-192 HDR source in order to investigate the influence of tissue heterogeneities on dose calculations compared to the dose in homogeneous water media, as it is typically calculated by brachytherapy Treatment Planning Systems (TPS). Validity of the simulation was verified in water medium in comparison with peer reviewed results using the dosimetric parameters recommended by AAPM, Task Group-43. The dose-rates in simulated prostate, bladder and rectum were compared to those obtained in the homogeneous water phantom. Based on the resulting dose differences, it is inferred that TPS algorithms for brachytherapy dose calculations overestimate the dose to tissues like prostate and bladder by up to 49%. A clinically relevant dose underestimation of 5.5% to the rectum was also found. We recommend that further investigation using actual patient CT data as input to the Monte Carlo simulation be performed. | |
Identifier: | 820353465 (oclc), 3355564 (digitool), FADT3355564 (IID), fau:3935 (fedora) | |
Note(s): |
by Ramsâes Herrera. Thesis (M.S.)--Florida Atlantic University, 2012. Includes bibliography. Mode of access: World Wide Web. System requirements: Adobe Reader. |
|
Subject(s): |
Medical physics Iridium -- Therapeutic use Imaging systems in medicine Radioisotope brachytherapy |
|