You are here
characterization of the LAP Aquarius Phantom for external LAP laser alignment and magnetic resonance geometric distortion verification for stereotactic radiation surgery patient simulation
- Date Issued:
- 2012
- Summary:
- The Thesis explores additional applications of LAP's Aquarius external laser alignment verification Phantom by examining geometric accuracy of magnetic resonance images commonly used for planning intracranial stereotactic radiation surgery (ICSRS) cases. The scans were performed with MRI protocols used for ICSRS, and head and neck diagnosis, and their images fused to computerized tomographic (CT) images. The geometric distortions (GDs) were measured against the CT in all axial, sagittal, and coronal directions at different levels. Using the Aquarius Phantom, one is able to detect GD in ICSRS planning MRI acquisitions, and align the external LAP patient alignment lasers, by following the LAP QA protocol. GDs up to about 2 mm are observed at the distal regions of the longitudinal axis in the SRS treatment planning MR images. Based on the results, one may recommend the use of the Aquarius Phantom to determine if margins should be included for SRS treatment planning.
Title: | A characterization of the LAP Aquarius Phantom for external LAP laser alignment and magnetic resonance geometric distortion verification for stereotactic radiation surgery patient simulation. |
438 views
53 downloads |
---|---|---|
Name(s): |
Vergara, Daniel Charles E. Schmidt College of Science Department of Physics |
|
Type of Resource: | text | |
Genre: | Electronic Thesis Or Dissertation | |
Date Issued: | 2012 | |
Publisher: | Florida Atlantic University | |
Physical Form: | electronic | |
Extent: | xvii, 73 p. : ill. (some col.) | |
Language(s): | English | |
Summary: | The Thesis explores additional applications of LAP's Aquarius external laser alignment verification Phantom by examining geometric accuracy of magnetic resonance images commonly used for planning intracranial stereotactic radiation surgery (ICSRS) cases. The scans were performed with MRI protocols used for ICSRS, and head and neck diagnosis, and their images fused to computerized tomographic (CT) images. The geometric distortions (GDs) were measured against the CT in all axial, sagittal, and coronal directions at different levels. Using the Aquarius Phantom, one is able to detect GD in ICSRS planning MRI acquisitions, and align the external LAP patient alignment lasers, by following the LAP QA protocol. GDs up to about 2 mm are observed at the distal regions of the longitudinal axis in the SRS treatment planning MR images. Based on the results, one may recommend the use of the Aquarius Phantom to determine if margins should be included for SRS treatment planning. | |
Identifier: | 810905734 (oclc), 3352878 (digitool), FADT3352878 (IID), fau:3914 (fedora) | |
Note(s): |
by Daniel Vergara. Thesis (M.S.)--Florida Atlantic University, 2012. Includes bibliography. Electronic reproduction. Boca Raton, Fla., 2012. Mode of access: World Wide Web. |
|
Subject(s): |
Radiotherapy -- Techniques Medical physics Imaging systems in medicine Tomographic images Computerized axial tomography |
|
Persistent Link to This Record: | http://purl.flvc.org/FAU/3352878 | |
Use and Reproduction: | http://rightsstatements.org/vocab/InC/1.0/ | |
Host Institution: | FAU |