Measurements of output factors for small photon fields up to 10 cm x 10 cm
- Date Issued:
- 2013
- Summary:
- Field output factors (OF) for photon beams from a 6 MV medical accelerator were measured using five different detectors in a scanning water phantom. The measurements were taken for square field sizes of integral widths ranging from 1 cm to 10 cm for two reference source-to-surface distances (SSD) and depths in water. For the diode detectors, square field widths as small as 2.5 mm were also studied. The photon beams were collimated by using either the jaws or the multileaf collimators. Measured OFs are found to depend upon the field size, SSD, depth and also upon the type of beam collimation, size and type of detector used. For field sizes larger than 3 cm x 3 cm, the OF measurements agree to within 1% or less. The largest variation in OF occurs for jawsshaped field of size 1 cm x 1cm, where a difference of more than 18% is observed.
Title: | Measurements of output factors for small photon fields up to 10 cm x 10 cm. |
123 views
55 downloads |
---|---|---|
Name(s): |
Bacala, Angelina, author Pella, Silvia, Thesis advisor Charles E. Schmidt College of Science, Degree grantor Department of Physics |
|
Type of Resource: | text | |
Genre: | Electronic Thesis Or Dissertation | |
Issuance: | single unit | |
Date Created: | Fall 2013 | |
Date Issued: | 2013 | |
Publisher: | Florida Atlantic University | |
Physical Form: | Online Resource | |
Extent: | 78 p. | |
Language(s): | English | |
Summary: | Field output factors (OF) for photon beams from a 6 MV medical accelerator were measured using five different detectors in a scanning water phantom. The measurements were taken for square field sizes of integral widths ranging from 1 cm to 10 cm for two reference source-to-surface distances (SSD) and depths in water. For the diode detectors, square field widths as small as 2.5 mm were also studied. The photon beams were collimated by using either the jaws or the multileaf collimators. Measured OFs are found to depend upon the field size, SSD, depth and also upon the type of beam collimation, size and type of detector used. For field sizes larger than 3 cm x 3 cm, the OF measurements agree to within 1% or less. The largest variation in OF occurs for jawsshaped field of size 1 cm x 1cm, where a difference of more than 18% is observed. | |
Identifier: | FA0004003 (IID) | |
Note(s): |
Includes bibliography. Thesis (M.S.)--Florida Atlantic University, 2013. |
|
Subject(s): |
Integrated circuits Photonics Quantum electrodynamics Quantum theory |
|
Held by: | Florida Atlantic University Digital Library | |
Sublocation: | Boca Raton, Fla. | |
Persistent Link to This Record: | http://purl.flvc.org/fau/fd/FA0004003 | |
Restrictions on Access: | All rights reserved by the source institution | |
Restrictions on Access: | http://rightsstatements.org/vocab/InC/1.0/ | |
Host Institution: | FAU |