An Algorithm to Determine IMRT Optimization Parameters within the Elekta Monaco® Treatment Planning System that Increases Dose Homogeneity and Dose Conformity in the Planning Target Volume

Download pdf | Full Screen View

Date Issued:
2022
Abstract/Description:
An algorithm to determine IMRT optimization parameters within the Elekta Monaco® treatment planning system that increases dose homogeneity and dose conformity in the planning target volume was developed. This algorithm determines IMRT optimization parameters by calculating the difference between two pairs of dose points along the target volume’s dose volume histogram: Dmax – Dmin, and D2 – D98. The algorithm was tested on the Elekta Monaco® Treatment Planning System at GenesisCare of Coconut Creek, Florida using CT data from 10 anonymized patients with non-small cell lung cancer of various tumor sizes and locations. Nine iterations of parameters were tested on each patient. Once the ideal parameters were found, the results were evaluated using the ICRU report 83 homogeneity index as well as the Paddick conformity index. As an outcome of this research, it is recommended that at least three iterations of IMRT optimization parameters should be calculated to find the ideal parameters.
Title: An Algorithm to Determine IMRT Optimization Parameters within the Elekta Monaco® Treatment Planning System that Increases Dose Homogeneity and Dose Conformity in the Planning Target Volume.
106 views
44 downloads
Name(s): Gregorisch, David , author
Pella, Silvia , Thesis advisor
Kyriacou, Andreas , Thesis advisor
Florida Atlantic University, Degree grantor
Department of Physics
Charles E. Schmidt College of Science
Type of Resource: text
Genre: Electronic Thesis Or Dissertation
Date Created: 2022
Date Issued: 2022
Publisher: Florida Atlantic University
Place of Publication: Boca Raton, Fla.
Physical Form: application/pdf
Extent: 125 p.
Language(s): English
Abstract/Description: An algorithm to determine IMRT optimization parameters within the Elekta Monaco® treatment planning system that increases dose homogeneity and dose conformity in the planning target volume was developed. This algorithm determines IMRT optimization parameters by calculating the difference between two pairs of dose points along the target volume’s dose volume histogram: Dmax – Dmin, and D2 – D98. The algorithm was tested on the Elekta Monaco® Treatment Planning System at GenesisCare of Coconut Creek, Florida using CT data from 10 anonymized patients with non-small cell lung cancer of various tumor sizes and locations. Nine iterations of parameters were tested on each patient. Once the ideal parameters were found, the results were evaluated using the ICRU report 83 homogeneity index as well as the Paddick conformity index. As an outcome of this research, it is recommended that at least three iterations of IMRT optimization parameters should be calculated to find the ideal parameters.
Identifier: FA00013990 (IID)
Degree granted: Thesis (PMS)--Florida Atlantic University, 2022.
Collection: FAU Electronic Theses and Dissertations Collection
Note(s): Includes bibliography.
Subject(s): Radiotherapy
Algorithm
Medical physics
Persistent Link to This Record: http://purl.flvc.org/fau/fd/FA00013990
Use and Reproduction: Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
Use and Reproduction: http://rightsstatements.org/vocab/InC/1.0/
Host Institution: FAU
Is Part of Series: Florida Atlantic University Digital Library Collections.