You are here
CONSISTENCY OF CT NUMBER AND ELECTRON DENSITY IN TREATMENT PLANNING SYSTEM VERSUS CT SCANNER, AND DOSIMETRIC CONSEQUENCES
- Date Issued:
- 2019
- Abstract/Description:
- The Computer Tomography (CT) scanned images are very important for the Treatment Planning System (TPS) to provide the electron density of the different types of tissues that the radiation penetrates in the path to the tumor to be treated. This electron density is converted to an attenuation coefficient, which varies with tissue for each structure and even varies by the tissue volume. The purpose of this research is to evaluate the CT numbers, and convert them into relative electron densities. Twenty-five patients’ data and CT numbers were evaluated in the CT scanner and in Eclipse and were converted into relative electron density and compared with each other. The differences between the relative electron density in the Eclipse was found to be from 0 up to 6% between tissue equivalent materials, the final result for all equivalent tissue materials was about 2%. For the patients’ data, the percentage difference of CT number versus electron density was found to be high for high relative electron density organs, namely the final average result for the spine was 8%, less for pelvis, and less for rib while for the other organs it was even less. The very lowest was 0.3% compared with 1% which is acceptable for clinical standards.
Title: | CONSISTENCY OF CT NUMBER AND ELECTRON DENSITY IN TREATMENT PLANNING SYSTEM VERSUS CT SCANNER, AND DOSIMETRIC CONSEQUENCES. |
158 views
72 downloads |
---|---|---|
Name(s): |
Hana, Evan Makdasy, author Pella, Silvia, Thesis advisor Leventouri, Theodora, Thesis advisor Florida Atlantic University, Degree grantor Charles E. Schmidt College of Science Department of Physics |
|
Type of Resource: | text | |
Genre: | Electronic Thesis Or Dissertation | |
Date Created: | 2019 | |
Date Issued: | 2019 | |
Publisher: | Florida Atlantic University | |
Place of Publication: | Boca Raton, Fla. | |
Physical Form: | application/pdf | |
Extent: | 107 p. | |
Language(s): | English | |
Abstract/Description: | The Computer Tomography (CT) scanned images are very important for the Treatment Planning System (TPS) to provide the electron density of the different types of tissues that the radiation penetrates in the path to the tumor to be treated. This electron density is converted to an attenuation coefficient, which varies with tissue for each structure and even varies by the tissue volume. The purpose of this research is to evaluate the CT numbers, and convert them into relative electron densities. Twenty-five patients’ data and CT numbers were evaluated in the CT scanner and in Eclipse and were converted into relative electron density and compared with each other. The differences between the relative electron density in the Eclipse was found to be from 0 up to 6% between tissue equivalent materials, the final result for all equivalent tissue materials was about 2%. For the patients’ data, the percentage difference of CT number versus electron density was found to be high for high relative electron density organs, namely the final average result for the spine was 8%, less for pelvis, and less for rib while for the other organs it was even less. The very lowest was 0.3% compared with 1% which is acceptable for clinical standards. | |
Identifier: | FA00013315 (IID) | |
Degree granted: | Thesis (M.S.)--Florida Atlantic University, 2019. | |
Collection: | FAU Electronic Theses and Dissertations Collection | |
Note(s): | Includes bibliography. | |
Subject(s): |
Tomography, X-Ray Computed Electron density Radiation dosimetry--Evaluation Tomography Scanners, X-Ray Computed |
|
Held by: | Florida Atlantic University Libraries | |
Sublocation: | Digital Library | |
Persistent Link to This Record: | http://purl.flvc.org/fau/fd/FA00013315 | |
Use and Reproduction: | Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder. | |
Use and Reproduction: | http://rightsstatements.org/vocab/InC/1.0/ | |
Host Institution: | FAU | |
Is Part of Series: | Florida Atlantic University Digital Library Collections. |