You are here
A dosimetric study of a heterogeneous phantom for lung stereotactic body radiation therapy comparing Monte Carlo and pencil beam calculations to dose distributions measured with a 2-d diode array
- Date Issued:
- 2015
- Abstract/Description:
- Monte Carlo (MC) and Pencil Beam (PB) calculations are compared to their measured planar dose distributions using a 2-D diode array for lung Stereotactic Body Radiation Therapy (SBRT). The planar dose distributions were studied for two different phantom types: an in-house heterogeneous phantom and a homogeneous phantom. The motivation is to mimic the human anatomy during a lung SBRT treatment and incorporate heterogeneities into the pre-treatment Quality Assurance process, where measured and calculated planar dose distributions are compared before the radiation treatment. Individual and combined field dosimetry has been performed for both fixed gantry angle (anterior to posterior) and planned gantry angle delivery. A gamma analysis has been performed for all beam arrangements. The measurements were obtained using the 2-D diode array MapCHECK 2™.
Title: | A dosimetric study of a heterogeneous phantom for lung stereotactic body radiation therapy comparing Monte Carlo and pencil beam calculations to dose distributions measured with a 2-d diode array. |
371 views
222 downloads |
---|---|---|
Name(s): |
Curley, Casey Michael, author Ouhib, Zoubir , Thesis advisor Florida Atlantic University, Degree grantor Charles E. Schmidt College of Science Department of Physics |
|
Type of Resource: | text | |
Genre: | Electronic Thesis Or Dissertation | |
Date Created: | 2015 | |
Date Issued: | 2015 | |
Publisher: | Florida Atlantic University | |
Place of Publication: | Boca Raton, Fla. | |
Physical Form: | application/pdf | |
Extent: | 139 p. | |
Language(s): | English | |
Abstract/Description: | Monte Carlo (MC) and Pencil Beam (PB) calculations are compared to their measured planar dose distributions using a 2-D diode array for lung Stereotactic Body Radiation Therapy (SBRT). The planar dose distributions were studied for two different phantom types: an in-house heterogeneous phantom and a homogeneous phantom. The motivation is to mimic the human anatomy during a lung SBRT treatment and incorporate heterogeneities into the pre-treatment Quality Assurance process, where measured and calculated planar dose distributions are compared before the radiation treatment. Individual and combined field dosimetry has been performed for both fixed gantry angle (anterior to posterior) and planned gantry angle delivery. A gamma analysis has been performed for all beam arrangements. The measurements were obtained using the 2-D diode array MapCHECK 2™. | |
Identifier: | FA00004360 (IID) | |
Degree granted: | Thesis (M.S.)--Florida Atlantic University, 2015. | |
Collection: | FAU Electronic Theses and Dissertations Collection | |
Note(s): | Includes bibliography. | |
Subject(s): |
Cancer -- Radiotherapy Drug development -- Computer simulation Image guided radiation therapy Ion bombardment Lung cancer -- Treatment Medical physics Monte Carlo method Proton beams |
|
Held by: | Florida Atlantic University Libraries | |
Sublocation: | Digital Library | |
Links: | http://purl.flvc.org/fau/fd/FA00004360 | |
Use and Reproduction: | Copyright © is held by the author, with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder. | |
Use and Reproduction: | http://rightsstatements.org/vocab/InC/1.0/ | |
Host Institution: | FAU | |